Every gauge drawing in IEC 60061-3 and every calibration certificate in the drawer assumes 20 °C. Most workshops are not at 20 °C in July. The gap is not a philosophical one: steel expands about 11.5 µm per metre per kelvin, so a 50 mm gauge used at 35 °C is roughly 8.6 µm larger than its certificate says.
Eight micrometres sounds small until it is compared with the tolerance it is checking, which for many lamp cap dimensions is 20 to 50 µm in total.
Expansion is linear: ΔL = α × L × ΔT. For steel, α is about 11.5 × 10-6 per K. At 15 K above reference, a 50 mm feature grows by 8.6 µm, a 100 mm feature by 17.3 µm, and a 200 mm feature by 34.5 µm.
| Feature size | At 25 °C | At 30 °C | At 35 °C |
| 10 mm | +0.6 µm | +1.2 µm | +1.7 µm |
| 50 mm | +2.9 µm | +5.8 µm | +8.6 µm |
| 100 mm | +5.8 µm | +11.5 µm | +17.3 µm |
| 200 mm | +11.5 µm | +23.0 µm | +34.5 µm |
ISO 1 fixes the reference temperature at 20 °C. It does not demand that every workshop be held at 20 °C; it demands that measurements be corrected to it, or that the deviation be shown to be negligible compared with the tolerance.
For a 10 mm feature with a 50 µm tolerance, 15 K of drift adds 1.7 µm, which is 3 % of the tolerance and can be ignored. For a 200 mm gauge with a 20 µm tolerance, the same drift adds 34.5 µm and the measurement means nothing.
Standards referenced: ISO 1 (standard reference temperature), ISO 14253-1 for uncertainty and decision rules.
Steel reaches thermal equilibrium slowly. A 50 mm gauge block needs about 40 minutes to settle after a 10 K change; a 200 mm gauge needs over two hours. Gauges brought in from a cold store or a van are not at room temperature when they arrive.
The practical rule is four hours on the bench before measuring, and overnight for anything above 100 mm. A gauge that feels room temperature to the hand is not necessarily at room temperature through its section.
It is not enough for both to be at 20 °C nominally. A part and a gauge that differ by 3 K produce a size error of about 1.7 µm on a 50 mm feature, which is the same order as the gauge tolerance.
Store parts and gauges in the same cabinet, not in different rooms, and handle parts with gloves so the hand does not warm one side.
If a measurement must be reported at 20 °C and the room is at 28 °C, the correction is subtractive for a steel part measured with a steel gauge: ΔL = 11.5 × 10
For a 100 mm feature that is 9.2 µm. Report the corrected value, the room temperature and the correction applied. A result without those three numbers cannot be compared with a result from another laboratory, which is usually the whole point of measuring.