A gauge is a comparator: its whole value comes from being a known size. That makes its own measurement the weakest link in the chain, and the two instruments used most often do not measure the same thing even when both read to a micrometre.
A three-point micrometer finds the maximum inscribed circle between three anvils. An air gauge measures the average clearance around a plug. On a perfect cylinder they agree; on a bore with a taper, an ovality or a burr, they diverge, and the divergence is exactly the size of the defect.
| Property | Three-point micrometer | Air gauge (pneumatic comparator) |
| Resolution | 1 µm typical | 0.1 µm on a good setting master |
| Measures | Maximum inscribed circle, at one depth | Average clearance over the plug length |
| Finds ovality | Yes, by rotating | No, it averages it away |
| Finds taper | Only by measuring at several depths | Directly, since it averages over length |
| Finds a burr | Usually reads it as a small bore | Often reads it as a large bore |
| Skill needed | Moderate | High, for setting and for masters |
| Best use | Final dimensional report | Production sorting of many parts |
Neither is wrong. A calibration report that uses one instrument and a shop floor that uses the other will produce a systematic difference, and that difference will be blamed on the gauge.
Both instruments are temperature-sensitive in the same way and for the same reason: steel moves about 11 µm per metre per kelvin. On a 10 mm bore that is 0.11 µm per kelvin, which sounds small until the gauge and the setting master sit on different benches.
A gauge certificate that lists a single diameter per feature is not comparable with the next one unless it also names the instrument, the number of readings and the temperature. Add the measurement uncertainty and the decision rule, and the certificate becomes something a second lab can reproduce.
The practical arrangement is a three-point micrometer for the periodic report and an air gauge for the daily check, with the two cross-checked on the same master every quarter. Our gauge and calibration accessories section covers setting masters and holders, and the wear planning that decides when a recheck happens is in the technical article list.
Standards referenced: ISO 14253-1 (decision rules), ISO/IEC 17025 clause 7.6 (uncertainty), ISO 3611 for micrometers.
Use the air gauge for the daily check and the micrometer for the certificate, and cross-check them on the same 10 mm master every quarter. If the two disagree by more than 2 µm, the air gauge setting master is the first suspect, not the gauge.
Write both numbers, with the instrument names, into the same record. An auditor who can see the pair does not need to ask how the decision was made.